Remeex Female

The Lifetime Readjustable Continence System

Unique lifetime readjustable sling

Personalized urethral support

Smallest sling

Excellent results in non easy cases

The Remeex Female is a system for SUI treatment that allows a reliable adjustment and re-adjustment of the sling to the desired urethral support whenever needed during patient’s lifetime. If used properly, surgeon can provide long-term continence avoiding retention.

Remeex Female provides the surgeon, the experience of adjusting the sling urethral support with the patient in standing up position and doing the efforts that drives her to incontinence. This unique property provides the surgeon the ability to adjust the sling to the desired urethral support level, avoiding urine leakage and urine retention.

Dynamic pressure transmission system

When coughing, laughing, etc. the patient advances the rectus abdominal muscle, advancing the Varitensor, and increasing the sling urethral support while it’s needed.

This reconstructs the intra-abdominal pressure transmission system of normal continent patients. The readjustment is made while the patient is standing up and doing the efforts that drives her to incontinence.

Lifetime readjustability

The sling adjustment is made step by step (checking continence every three turns of the manipulator). When the patient is continent we know that three turns before she was incontinent.

This allows the surgeon to provide the minimum necessary urethral support for that particular patient under stress and no more, to achieve continence and avoid retention.

This system has three simple steps:

1. Surgery: The Remeex Female is placed at the OR leaving the system totally tension free (patient still incontinent).

2. Post-op Adjustment: 24-48h after the intervention the system is adjusted up to the desired urethral support level. This adjustment is done with the patient standing up and doing the movements that drives him to incontinence. By performing this adjustment properly step by step, continence can be achieved at the right level avoiding retention.

3. Re-Adjustment: The system can be re-adjusted as many times as needed during patient’s lifetime. When needed, surgeon can adjust the sling urethral support by reconnecting the manipulator to the varitensor to increase the urethral support level up to continence.